Mean surface roughness (Ra, Rp and Rv) of cobalt chrome and PEEK-OPTIMA™ femoral components (n=3) after 3 million cycles testing under intermediate kinematics and 3 million cycles of testing under high kinematic conditions. Each value is the average of 4 measurements taken within the flexion range of the study. Data for intermediate kinematics presented in Table 1, data for high kinematics presented in Table 3 of the manuscript. INTERMEDIATE KINEMATICS Ra cobalt chrome PEEK-OPTIMA™ SAMPLE Pre-test Post-test Pre-test Post-test 1 0.024 0.052 0.018 0.274 2 0.023 0.024 0.018 0.185 3 0.023 0.024 0.019 0.143 Rp cobalt chrome PEEK-OPTIMA™ SAMPLE Pre-test Post-test Pre-test Post-test 1 0.082 0.139 0.077 0.597 2 0.082 0.080 0.076 0.413 3 0.080 0.082 0.081 0.285 Rv cobalt chrome PEEK-OPTIMA™ SAMPLE Pre-test Post-test Pre Post-test 1 0.063 0.139 0.067 1.233 2 0.062 0.071 0.065 1.055 3 0.058 0.066 0.073 0.877 HIGH KINEMATICS Ra cobalt chrome PEEK-OPTIMA™ SAMPLE Pre-test Post-test Pre-test Post-test 1 0.052 0.032 0.274 0.252 2 0.024 0.033 0.185 0.195 3 0.024 0.026 0.143 0.151 Rp cobalt chrome PEEK-OPTIMA™ SAMPLE Pre-test Post-test Pre-test Post-test 1 0.139 0.093 0.597 0.586 2 0.080 0.096 0.413 0.381 3 0.082 0.079 0.285 0.364 Rv cobalt chrome PEEK-OPTIMA™ SAMPLE Pre-test Post-test Pre Post-test 1 0.139 0.097 1.233 0.765 2 0.071 0.104 1.055 0.537 3 0.066 0.080 0.877 0.513