New TopShim run:

version = topshim 2.0.0
date = 2023-11-14
time = 10:29:36.586
command = topshim  convcomp tuneaxyz

1D SHIMMING

Parameters:

maximum order = 5
probe = Z116098_0395
solvent = CDCl3
shim nucleus = 2H
nucleus optimised for = 1H
o1p (from lock) = 7.24 ppm
optimisation parameters = ls
linewidth 1H = 0.10 Hz
envelope shape / strictness = 2.00 / 200

Results:

tune lock level increase = 1.5

initial B0 stdDev = 3.11 Hz
sample size = 2.85 cm, position = 0.02 cm
final B0 stdDev = 0.33 Hz > improvement = 9.5
envelope width = 0.73 Hz

Performing tune after shimming
tune lock level increase = 1.0

shim changes: 
Z	+77
Z2	-40
Z3	+79
Z4	-270
Z5	+111
X	-53
Y	+125

duration = 3 min 48 sec
completed successfully
finished Tue Nov 14 10:33:23 2023


========================================================================