New TopShim run:

version = topshim 2.0.0
date = 2023-11-14
time = 18:14:06.749
command = topshim  convcomp tuneaxyz

1D SHIMMING

Parameters:

maximum order = 5
probe = Z116098_0395
solvent = CDCl3
shim nucleus = 2H
nucleus optimised for = 13C
o1p (from lock) = 7.24 ppm
optimisation parameters = ls
linewidth 1H = 0.10 Hz
envelope shape / strictness = 2.00 / 200

Results:

tune lock level increase = 1.4

initial B0 stdDev = 3.04 Hz
sample size = 2.86 cm, position = 0.00 cm
final B0 stdDev = 0.34 Hz > improvement = 8.9
envelope width = 0.74 Hz

Performing tune after shimming
tune lock level increase = 1.0

shim changes: 
Z	+62
Z2	-39
Z3	+116
Z4	-272
Z5	+70
X	-38
Y	+131

duration = 3 min 37 sec
completed successfully
finished Tue Nov 14 18:17:42 2023


========================================================================