New TopShim run:

version = topshim 2.0.0
date = 2023-11-15
time = 12:15:25.064
command = topshim  convcomp tuneaxyz

1D SHIMMING

Parameters:

maximum order = 5
probe = Z116098_0395
solvent = CDCl3
shim nucleus = 2H
nucleus optimised for = 1H
o1p (from lock) = 7.24 ppm
optimisation parameters = ls
linewidth 1H = 0.10 Hz
envelope shape / strictness = 2.00 / 200

Results:

tune lock level increase = 1.4

initial B0 stdDev = 2.96 Hz
sample size = 2.90 cm, position = 0.01 cm
final B0 stdDev = 0.40 Hz > improvement = 7.5
envelope width = 0.69 Hz

Performing tune after shimming
tune lock level increase = 1.0

shim changes: 
Z	+53
Z2	-39
Z3	+128
Z4	-235
Z5	-19
X	-59
Y	+138

duration = 3 min 15 sec
completed successfully
finished Wed Nov 15 12:18:38 2023


========================================================================