New TopShim run:

version = topshim 2.0.0
date = 2023-11-15
time = 18:07:13.331
command = topshim  convcomp tuneaxyz

1D SHIMMING

Parameters:

maximum order = 5
probe = Z116098_0395
solvent = CDCl3
shim nucleus = 2H
nucleus optimised for = 13C
o1p (from lock) = 7.24 ppm
optimisation parameters = ls
linewidth 1H = 0.10 Hz
envelope shape / strictness = 2.00 / 200

Results:

tune lock level increase = 1.4

initial B0 stdDev = 2.68 Hz
sample size = 2.88 cm, position = -0.01 cm
final B0 stdDev = 0.32 Hz > improvement = 8.4
envelope width = 0.66 Hz

Performing tune after shimming
tune lock level increase = 1.0

shim changes: 
Z	+57
Z2	-38
Z3	+93
Z4	-219
Z5	+40
X	-48
Y	+121

duration = 3 min 51 sec
completed successfully
finished Wed Nov 15 18:11:03 2023


========================================================================