New TopShim run:

version = topshim 2.0.0
date = 2023-11-16
time = 10:53:13.525
command = topshim  convcomp tuneaxyz

1D SHIMMING

Parameters:

maximum order = 5
probe = Z116098_0395
solvent = CDCl3
shim nucleus = 2H
nucleus optimised for = 1H
o1p (from lock) = 7.24 ppm
optimisation parameters = ls
linewidth 1H = 0.10 Hz
envelope shape / strictness = 2.00 / 200

Results:

tune lock level increase = 1.2

initial B0 stdDev = 3.38 Hz
sample size = 2.72 cm, position = 0.08 cm
final B0 stdDev = 0.40 Hz > improvement = 8.5
envelope width = 1.39 Hz

Performing tune after shimming
tune lock level increase = 1.0

shim changes: 
Z	+86
Z2	-41
Z3	+63
Z4	-310
Z5	+224
X	-36
Y	+125

duration = 3 min 20 sec
completed successfully
finished Thu Nov 16 10:56:32 2023


========================================================================