New TopShim run:

version = topshim 2.0.0
date = 2023-11-15
time = 12:05:05.407
command = topshim  convcomp tuneaxyz

1D SHIMMING

Parameters:

maximum order = 5
probe = Z116098_0395
solvent = CDCl3
shim nucleus = 2H
nucleus optimised for = 1H
o1p (from lock) = 7.24 ppm
optimisation parameters = ls
linewidth 1H = 0.10 Hz
envelope shape / strictness = 2.00 / 200

Results:

tune lock level increase = 1.4

initial B0 stdDev = 2.93 Hz
sample size = 2.85 cm, position = 0.00 cm
final B0 stdDev = 0.32 Hz > improvement = 9.1
envelope width = 0.69 Hz

Performing tune after shimming
tune lock level increase = 1.0

shim changes: 
Z	+65
Z2	-40
Z3	+102
Z4	-240
Z5	+53
X	-48
Y	+123

duration = 3 min 51 sec
completed successfully
finished Wed Nov 15 12:08:55 2023


========================================================================