New TopShim run:

version = topshim 2.0.0
date = 2023-11-16
time = 19:27:35.424
command = topshim  convcomp tuneaxyz

1D SHIMMING

Parameters:

maximum order = 5
probe = Z116098_0395
solvent = CDCl3
shim nucleus = 2H
nucleus optimised for = 13C
o1p (from lock) = 7.24 ppm
optimisation parameters = ls
linewidth 1H = 0.10 Hz
envelope shape / strictness = 2.00 / 200

Results:

tune lock level increase = 1.3

initial B0 stdDev = 2.87 Hz
sample size = 2.84 cm, position = 0.01 cm
final B0 stdDev = 0.36 Hz > improvement = 8.0
envelope width = 0.98 Hz

Performing tune after shimming
tune lock level increase = 1.0

shim changes: 
Z	+58
Z2	-37
Z3	+117
Z4	-248
Z5	+47
X	-45
Y	+132

duration = 3 min 48 sec
completed successfully
finished Thu Nov 16 19:31:22 2023


========================================================================